Drug-coated balloon versus uncoated percutaneous transluminal angioplasty for the treatment of atherosclerotic lesions in the superficial femoral and proximal popliteal artery: 2-year results of the MDT-2113 SFA Japan randomized trial

Catheter Cardiovasc Interv. 2019 Mar 1;93(4):664-672. doi: 10.1002/ccd.28048. Epub 2019 Feb 12.

Abstract

Objectives: To assess the longer-term safety and efficacy of the IN.PACT Admiral (MDT-2113) drug-coated balloon (DCB) for the treatment of de novo and non-stented restenotic lesions in the superficial femoral and/or proximal popliteal arteries versus uncoated percutaneous transluminal angioplasty (PTA) in a Japanese cohort.

Background: Although DCBs are the newest endovascular strategy for patients with peripheral artery disease presenting with femoropopliteal lesions, there remains a paucity of results in non-Caucasian populations.

Methods: IN.PACT SFA Japan is an independently-adjudicated, prospective, multicenter, randomized, single-blinded trial. Endpoints through 2 years included primary patency and a composite safety endpoint of freedom from device- and procedure-related death through 30 days, freedom from target limb major amputation and freedom from clinically-driven target vessel revascularization at 24 months.

Results: One hundred patients were assigned by 2:1 randomization to treatment with the IN.PACT Admiral DCB (n = 68) or PTA (n = 32). The groups were well-matched at baseline. Mean lesion length for the DCB and PTA groups were 9.15 ± 5.85 and 8.89 ± 6.01 cm (P = 0.838), respectively. Patients treated with DCB exhibited superior 24-month primary patency compared to PTA (79.8% vs. 46.9%; log rank P < 0.001). The 24-month clinically driven target lesion revascularization rate was 9.1% for DCB versus 20.7% for PTA (P = 0.177). There were no device- or procedure-related deaths, major amputations, or thromboses in either group.

Conclusions: Two-year results from IN.PACT SFA Japan demonstrated persistently superior patency and low CD-TLR rates through 2 years when compared to uncoated PTA in Japanese patients. These data are consistent with other IN.PACT DCB trials.

Keywords: balloon angioplasty; drug-coated balloon; peripheral artery disease.

Publication types

  • Clinical Trial, Phase III
  • Comparative Study
  • Multicenter Study
  • Randomized Controlled Trial

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Amputation, Surgical
  • Angioplasty, Balloon / adverse effects
  • Angioplasty, Balloon / instrumentation*
  • Cardiovascular Agents / administration & dosage*
  • Cardiovascular Agents / adverse effects
  • Coated Materials, Biocompatible*
  • Equipment Design
  • Female
  • Femoral Artery* / diagnostic imaging
  • Femoral Artery* / physiopathology
  • Humans
  • Japan
  • Limb Salvage
  • Male
  • Middle Aged
  • Peripheral Arterial Disease / diagnostic imaging
  • Peripheral Arterial Disease / physiopathology
  • Peripheral Arterial Disease / therapy*
  • Popliteal Artery* / diagnostic imaging
  • Popliteal Artery* / physiopathology
  • Progression-Free Survival
  • Prospective Studies
  • Risk Factors
  • Single-Blind Method
  • Time Factors
  • Vascular Access Devices*
  • Vascular Patency
  • Young Adult

Substances

  • Cardiovascular Agents
  • Coated Materials, Biocompatible