Multipoint left ventricular pacing improves response to cardiac resynchronization therapy with and without pressure-volume loop optimization: comparison of the long-term efficacy of two different programming strategies

J Interv Card Electrophysiol. 2019 Mar;54(2):141-149. doi: 10.1007/s10840-018-0480-6. Epub 2018 Nov 27.

Abstract

Purpose: Cardiac resynchronization therapy (CRT) with multipoint left ventricular (LV) pacing (MultiPoint™ Pacing [MPP]) improves long-term LV reverse remodeling, though questions persist about how to program LV pacing vectors and delays. We evaluated if an empirical method of programming MPP vectors and delays between pacing pulses improved CRT response similar to pressure-volume loop (PVL) optimized MPP programming.

Methods: Patients undergoing CRT implant (Quadra Assura MP™ CRT-D and Quartet™ LV lead) received MPP with programmed settings optimized either by PVL measurements at implant (PVL-OPT group) or empirically determined by maximizing the spatial separation between the two cathodes and minimal delays between the three ventricular pacing pulses (MAX-SEP group). CRT response was prospectively defined as a reduction in end-systolic volume (ESV) of ≥ 15% relative to baseline at 6 months as determined by a blinded observer.

Results: Patient characteristics at baseline (NYHA II-III, ejection fraction [EF] 27 ± 6%, QRS 151 ± 17 ms) were not significantly different between the PVL-OPT (n = 27) and MAX-SEP (n = 26) groups. During the follow-up period, there were no differences in the number of patients requiring reprogramming due to phrenic nerve stimulation or a high threshold for PVL-OPT vs. MAX-SEP (5/27 [19%] vs. 7/26 [27%], p = 0.53). After 6 months, ESV reduction, EF increase, and CRT response rate (RR) were similar for PVL-OPT vs. MAX-SEP (ESV - 20 ± 11 vs. - 22 ± 11%, p = 0.59; EF + 10 ± 4 vs. + 9 ± 7%, p = 0.53; RR 20/27 [74%] vs. 21/26 [81%], p = 0.74), while fewer patients in the PVL-OPT group experienced NYHA class reduction ≥ 2 (4/27 [15%] vs.15/26 [58%], p = 0.002).

Conclusions: Both evaluated methods of MPP programming resulted in similar CRT outcomes. Empirical MPP programming by maximum spatial separation of LV cathodes may be an effective, simple, and non-invasive alternative to pressure-volume optimization.

Keywords: CRT response; Cardiac resynchronization therapy; Heart failure; Multipoint pacing.

Publication types

  • Comparative Study

MeSH terms

  • Aged
  • Cardiac Pacing, Artificial / methods*
  • Cardiac Resynchronization Therapy / methods
  • Cohort Studies
  • Echocardiography / methods*
  • Female
  • Heart Failure / diagnosis
  • Heart Failure / therapy*
  • Heart Ventricles / diagnostic imaging*
  • Humans
  • Male
  • Middle Aged
  • Prognosis
  • Prospective Studies
  • Risk Assessment
  • Stroke Volume / physiology
  • Time Factors
  • Treatment Outcome
  • Ventricular Remodeling / physiology*