Limitations of platform assays to measure serum 25OHD level impact on guidelines and practice decision making

Metabolism. 2018 Dec:89:1-7. doi: 10.1016/j.metabol.2018.09.003. Epub 2018 Sep 15.

Abstract

Context: Liquid Chromatography Mass Spectroscopy (LC-MS/MS) is the preferred method to measure 25 hydroxyvitamin D (25OHD) levels, but laboratories are increasingly adopting automated platform assays.

Objective: We assessed the performance of commonly used automated immunoassays, with that of LC-MS/MS, and the National Institute of Standards and Technology (NIST) reference values, to measure 25OHD levels.

Methods/setting: We compared serum 25OHD levels obtained from 219 elderly subjects, enrolled in a vitamin D trial, using the Diasorin Liaison platform assay, and the tandem LC-MS/MS method. We also assessed the performance of the Diasorin and Roche automated assays, expressed as mean % bias from the NIST standards, based on the vitamin D External Quality Assessment Scheme (DEQAS) reports, from 2013 to 2017.

Results: Serum 25OHD levels were significantly lower in the Diasorin compared to LC-MS/MS assay at baseline, 18.5 ± 7.8 vs 20.5 ± 7.6 ng/ml (p < 0.001), and all other time points. Diasorin (25OHD) = 0.76 × LC-MS/MS (25OHD) + 4.3, R2 = 0.596. The absolute bias was independent of 25OHD values, and the pattern unfit for any cross-calibration. The proportion of subjects considered for vitamin D treatment based on pre-set cut-offs differed significantly between the 2 assays. There also was wide variability in the performance of both automated assays, compared to NIST reference values.

Conclusion: The performance of most widely used automated assays is sub-optimal. Our findings underscore the pressing need to re-consider current practices with regard to 25OHD measurements, interpretation of results from research studies, meta-analyses, the development of vitamin D guidelines, and their relevance to optimizing health.

Trial registration: ClinicalTrials.gov NCT01315366.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Automation
  • Chromatography, High Pressure Liquid
  • Clinical Decision-Making*
  • Female
  • Guidelines as Topic
  • Humans
  • Immunoassay
  • Luminescence
  • Male
  • Mass Spectrometry
  • Nutritional Status
  • Overweight / blood
  • Reference Values
  • Reproducibility of Results
  • Vitamin D / analogs & derivatives*
  • Vitamin D / blood
  • Vitamin D Deficiency / blood*

Substances

  • Vitamin D
  • 25-hydroxyvitamin D

Associated data

  • ClinicalTrials.gov/NCT01315366