Background: Metal electrodes are a mainstay of neuroscience. Characterization of the electrical impedance properties of these cuffs is important to ensure successful and repeatable fabrication, achieve a target impedance, revise novel designs, and quantify the success or failure of implantation and any potential subsequent damage or encapsulation by scar tissue.
New methods: Impedances are frequently characterized using lumped-parameter circuit models of the electrode-electrolyte interface. Open-source tools to gather and analyze these frequency sweep data are lacking. Here, we present such software, in the form of Matlab code, which includes a GUI. It automatically acquires frequency sweep data and subsequently fits a simplified Randles model to these data, over a user specified frequency range, providing the user with the model parameter estimates. Also, it can measure an unknown impedance of an element over a range of frequencies, as long as an external resistor can be added for the measurements.
Results: The tool was tested on five bright platinum nerve cuffs in vitro. The average charge transfer resistance, solution resistance, CPE value, and impedance magnitude were estimated.
Comparison to existing methods: The measured values of the impedance of cuffs were in agreement with the literature (Wei and Grill, 2009). Variation between cuffs fabricated as consistently as possible amounted to 10% for impedance magnitude and 4° for impedance phase.
Conclusion: The results show that this low-cost tool can be used to characterize a cuff across different conditions including after implantation. The latter makes it useful for a longer-term study of electrode viability.
Keywords: Impedance measurement; Matlab code; Nerve cuffs; Neuroscience; Signal processing; Software.