Incidence, Predictors, and Significance of Ventricular Arrhythmias in Patients With Continuous-Flow Left Ventricular Assist Devices: A 15-Year Institutional Experience

JACC Clin Electrophysiol. 2018 Feb;4(2):257-264. doi: 10.1016/j.jacep.2017.11.001. Epub 2017 Dec 20.

Abstract

Objectives: The aim of this study was to evaluate the incidence, predictors, and associated mortality of pre-implantation, early, and late ventricular arrhythmias (VAs) in patients receiving continuous-flow left ventricular assist devices (CFLVADs).

Background: VAs are common both pre- and post-implantation of left ventricular assist devices. Limited data exist on their prognostic impact in contemporary CFLVADs.

Methods: A retrospective review was performed to identify patients who underwent CFLVAD implantation between 2000 and 2015 with 2 years of follow-up. All VAs, defined as ventricular fibrillation, ventricular tachycardia lasting >30 s, or a ventricular rhythm requiring defibrillation, were analyzed. VAs occurring within 30 days of implantation were defined as early. Recorded outcomes included death and receipt of cardiac transplant.

Results: A total of 517 patients were included for analysis. Early VAs were associated with a significant reduction in survival (hazard ratio: 1.83; 95% confidence interval: 1.28 to 2.61; p = 0.001) compared with patients with late or no VAs. Pre-implantation variables independently predictive of early VAs included prior cardiac surgery (odds ratio: 1.90; 95% confidence interval: 1.09 to 3.32; p = 0.023) and pre-CFLVAD ventricular tachycardia storm (odds ratio: 3.15; 95% confidence interval: 1.49 to 6.69; p = 0.003). The incidence of early VAs from 2000 to 2007 was as high as 47%, whereas the highest incidence from 2008 to 2015 was <22%.

Conclusions: VAs within 30 days after CFLVAD implantation are associated with an increased risk for death. Predictors of early VAs include prior cardiac surgery and pre-CFLVAD ventricular tachycardia storm. Temporal trends have shown a decrease in VA from 2000 to 2015. Strategies to reduce arrhythmia burden shortly after CFLVAD implantation warrant further investigation.

Keywords: cardiac arrhythmia; continuous-flow left ventricular assist device; left ventricular assist device; ventricular arrhythmia; ventricular fibrillation; ventricular tachycardia.

MeSH terms

  • Aged
  • Arrhythmias, Cardiac* / epidemiology
  • Arrhythmias, Cardiac* / mortality
  • Female
  • Heart Ventricles / physiopathology
  • Heart-Assist Devices* / adverse effects
  • Heart-Assist Devices* / statistics & numerical data
  • Humans
  • Incidence
  • Kaplan-Meier Estimate
  • Male
  • Middle Aged
  • Retrospective Studies
  • Risk Factors