The current iterations of balloon-expandable transcatheter aortic valve replacement (TAVR) valves consist of a maximum size of 29 mm, corresponding to maximum annulus area of 680 mm2. However, a number of patients who qualify for TAVR based on surgical risk may have anatomical features outside the FDA-approved descriptions. The technique of overexpansion of TAVR valves during deployment allows for more patients to be treated successfully. This particular case demonstrates that overexpansion is safe and efficacious in a very large annulus of >800 mm2. Careful planning and consideration of aortic root calcification is helpful for technique of overexpansion and, ultimately, the technique may increase the number of patients eligible for TAVR.
Keywords: Annulus; Aortic stenosis; Bicuspid aortic valves; Surgical aortic valve replacement; Transcatheter aortic valve replacement.
Copyright © 2018. Published by Elsevier Inc.