Objectives: Imported Chagas disease (CD) is an emerging health problem in Europe due to immigration from endemic countries. Although WHO currently recommends two different serological methods to establish diagnosis, new tools like the ARCHITECT Chagas assay have potential for use as a single diagnostic test. Our objective was to determine an optimal signal-to-cut-off (S/CO) value for the ARCHITECT Chagas assay to diagnose CD with a single test.
Methods: A retrospective study conducted at the 12 de Octubre University Hospital (Madrid, Spain). All patients with requests for Chagas screening between January 2014 and August 2017 were consecutively included. All samples were routinely tested with the ARCHITECT assay. Negative samples (S/CO < 0.8) required no further testing. Immunochromatographic testing (ICT) and/or indirect immunofluorescence (IFI) was used to confirm samples with S/CO ≥ 0.8. Receiver operator characteristic (ROC) curve analysis determined the ARCHITECT S/CO value that yielded 100% specificity and positive predictive value. SPSS software, version 22.0 was used for data analysis.
Results: A total of 4153 samples were analysed; 361 (8.69%) gave a reactive ARCHITECT Chagas result. 261/361 (72.3%) were women; median age was 38 years old (2-79). 92.8% were Bolivian. A total of 307 (85.0%) were confirmed as cases of Chagas; 52 (14.4%) were not infected; two (0.6%) were not evaluable. Seroprevalence was 7.39%. An S/CO ≥ 3.80 yielded 100% specificity (95% confidence interval [CI], 0.93-1.00) and 100% positive predictive value (95% CI, 0.99-1.00).
Conclusions: Using S/CO ≥ 3.80, the ARCHITECT Chagas could be used as a single test for diagnosis of chronic CD in Bolivian immigrants. Patients with S/CO between 0.80 and 3.80 would require additional testing.
Keywords: ARCHITECT Chagas assay; Chagas disease; cut-off value; diagnostic; diagnostics; maladie de Chagas; test Architecte Chagas; valeur seuil.
© 2018 John Wiley & Sons Ltd.