Ex vivo heart perfusion has been shown to be an effective means of facilitating the resuscitation and assessment of donor hearts for cardiac transplantation. Over the last ten years however, only a few ex vivo perfusion systems have been developed for this application. While results have been promising, a system capable of facilitating multiple perfusion strategies on the same platform has not yet been realized. In this paper, the design, development and testing of a novel and modular ex vivo perfusion system is described. The system is capable of operating in three unique primary modes: the traditional Langendorff Mode, Pump-Supported Working-Mode, and Passive Afterload Working-Mode. In each mode, physiological hemodynamic parameters can be produced by managing perfusion settings. To evaluate heart viability, six experiments were conducted using porcine hearts and measuring several parameters including: pH, aortic pressure, lactate metabolism, coronary vascular resistance (CVR), and myocardial oxygen consumption. Pressure-volume relationship measurements were used to assess left ventricular contractility in each Working Mode. Hemodynamic and metabolic conditions remained stable and consistent across 4 h of ex vivo heart perfusion on the ex vivo perfusion system, validating the system as a viable platform for future development of novel preservation and assessment strategies.
Keywords: Biomechanics; Cardiac; Ex vivo; Heart transplantation; Perfusion system.