Three-Year Outcomes With the Absorb Bioresorbable Scaffold: Individual-Patient-Data Meta-Analysis From the ABSORB Randomized Trials

Circulation. 2018 Jan 30;137(5):464-479. doi: 10.1161/CIRCULATIONAHA.117.031843. Epub 2017 Oct 31.

Abstract

Background: The Absorb bioresorbable vascular scaffold (BVS) completely resorbs within 3 years after coronary artery implantation. The safety and effectiveness of BVS through this critical 3-year period have not been characterized.

Methods: We performed an individual-patient-data pooled meta-analysis of the 4 randomized ABSORB trials in which 3389 patients with coronary artery disease were randomly assigned to everolimus-eluting Absorb BVS (n=2164) or cobalt-chromium everolimus-eluting stents (n=1225). The primary efficacy outcome measure was target lesion failure (cardiac mortality, target vessel myocardial infarction, or ischemia-driven target lesion revascularization), and the primary safety outcome measure was device thrombosis.

Results: BVS compared with cobalt-chromium everolimus-eluting stents resulted in higher 3-year rates of target lesion failure (11.7% versus 8.1%; risk ratio [RR], 1.38; 95% confidence interval [CI], 1.10-1.73; P=0.006), driven by greater target vessel myocardial infarction (7.8% versus 4.2%; RR, 1.72; 95% CI, 1.26-2.35; P=0.0006) and ischemia-driven target lesion revascularization (6.6% versus 4.4%; RR, 1.44; 95% CI, 1.05-1.98; P=0.02), with comparable cardiac mortality (1.1% versus 1.1%; RR, 0.93; 95% CI, 0.47-1.88; P=0.85). Device thrombosis rates through 3 years were also higher with BVS (2.4% versus 0.6%; RR, 3.71; 95% CI, 1.70-8.11; P=0.001). Between 1 and 3 years, target lesion failure rates (6.1% versus 3.9%; P=0.02) and device thrombosis rates (1.1% versus 0.0%; P<0.0001) were higher with BVS than cobalt-chromium everolimus-eluting stents.

Conclusions: In the present individual-patient-data pooled meta-analysis of the ABSORB trials, BVS was associated with increased rates of target lesion failure and device thrombosis between 1 and 3 years and cumulatively through 3 years of follow-up compared with everolimus-eluting stents.

Clinical trial registration: URL: https://clinicaltrials.gov. Unique identifiers: NCT01751906, NCT01844284, NCT01923740, and NCT01425281.

Keywords: coronary artery disease; drug-eluting stents; meta-analysis [publication type]; percutaneous coronary intervention; stents.

Publication types

  • Meta-Analysis
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Absorbable Implants*
  • Coronary Artery Disease / diagnostic imaging
  • Coronary Artery Disease / mortality
  • Coronary Artery Disease / surgery*
  • Coronary Thrombosis / epidemiology
  • Drug-Eluting Stents*
  • Humans
  • Myocardial Infarction / diagnostic imaging
  • Myocardial Infarction / mortality
  • Myocardial Infarction / surgery*
  • Percutaneous Coronary Intervention / adverse effects
  • Percutaneous Coronary Intervention / instrumentation*
  • Percutaneous Coronary Intervention / mortality
  • Prosthesis Design
  • Randomized Controlled Trials as Topic
  • Recurrence
  • Risk Factors
  • Time Factors
  • Treatment Outcome

Associated data

  • ClinicalTrials.gov/NCT01751906
  • ClinicalTrials.gov/NCT01844284
  • ClinicalTrials.gov/NCT01923740
  • ClinicalTrials.gov/NCT01425281