Pulmonary Drug Delivery Following Continuous Vibrating Mesh Nebulization and Inspiratory Synchronized Vibrating Mesh Nebulization During Noninvasive Ventilation in Healthy Volunteers

J Aerosol Med Pulm Drug Deliv. 2018 Feb;31(1):33-41. doi: 10.1089/jamp.2016.1339. Epub 2017 Jul 6.

Abstract

Background: A breath-synchronized nebulization option that could potentially improve drug delivery during noninvasive positive pressure ventilation (NIPPV) is currently not available on single-limb circuit bilevel ventilators. The aim of this study was to compare urinary excretion of amikacin following aerosol delivery with a vibrating mesh nebulizer coupled to a single-limb circuit bilevel ventilator, using conventional continuous (Conti-Neb) and experimental inspiratory synchronized (Inspi-Neb) nebulization modes.

Materials and methods: A crossover clinical trial involving 6 noninvasive ventilated healthy volunteers (mean age of 32.3 ± 9.5 y) randomly assigned to both vibrating mesh nebulization modes was conducted: Inspi-Neb delivered aerosol during only the whole inspiratory phase, whereas Conti-Neb delivered aerosol continuously. All subjects inhaled amikacin solution (500 mg/4 mL) during NIPPV using a single-limb bilevel ventilator (inspiratory positive airway pressure: 12 cm H2O, and expiratory positive airway pressure: 5 cm H2O). Pulmonary drug delivery of amikacin following both nebulization modes was compared by urinary excretion of drug for 24 hours post-inhalation.

Results: The total daily amount of amikacin excreted in the urine was significantly higher with Inspi-Neb (median: 44.72 mg; interquartile range [IQR]: 40.50-65.13) than with Conti-Neb (median: 40.07 mg; IQR: 31.00-43.73), (p = 0.02). The elimination rate constant of amikacin (indirect measure of the depth of drug penetration into the lungs) was significantly higher with Inspi-Neb (median: 0.137; IQR: 0.113-0.146) than with Conti-Neb (median: 0.116; IQR: 0.105-0.130), (p = 0.02). However, the mean pulmonary drug delivery rate, expressed as the ratio between total daily urinary amount of amikacin and nebulization time, was significantly higher with Conti-Neb (2.03 mg/min) than with Inspi-Neb (1.09 mg/min) (p < 0.01).

Conclusions: During NIPPV with a single-limb circuit bilevel ventilator, the use of inspiratory synchronized vibrating mesh nebulization may improve pulmonary drug delivery compared with conventional continuous vibrating mesh nebulization.

Keywords: aerosol delivery; amikacin; bilevel ventilator; breath-synchronized nebulizer; healthy volunteers; nebulization; nebulization mode; noninvasive positive pressure ventilation; pulmonary drug delivery; single-limb circuit ventilator; vibrating mesh nebulizer.

Publication types

  • Comparative Study
  • Randomized Controlled Trial

MeSH terms

  • Administration, Inhalation
  • Adult
  • Aerosols
  • Amikacin / administration & dosage*
  • Amikacin / pharmacokinetics
  • Anti-Bacterial Agents / administration & dosage*
  • Anti-Bacterial Agents / pharmacokinetics
  • Cross-Over Studies
  • Double-Blind Method
  • Drug Delivery Systems*
  • Humans
  • Lung / metabolism
  • Male
  • Middle Aged
  • Nebulizers and Vaporizers
  • Noninvasive Ventilation*
  • Tissue Distribution
  • Vibration
  • Young Adult

Substances

  • Aerosols
  • Anti-Bacterial Agents
  • Amikacin