Precision of pQCT-measured total, trabecular and cortical bone area, content, density and estimated bone strength in children

J Musculoskelet Neuronal Interact. 2017 Jun 1;17(2):59-68.

Abstract

Objectives: To define pQCT precision errors, least-significant-changes, and identify associated factors for bone outcomes at the radius and tibia in children.

Methods: We obtained duplicate radius and tibia pQCT scans from 35 children (8-14yrs). We report root-mean-squared coefficient of variation (CV%RMS) and 95% limits-of-agreement to characterize repeatability across scan quality and least-significant-changes for bone outcomes at distal (total and trabecular area, content and density; and compressive bone strength) and shaft sites (total area and content; cortical area content, density and thickness; and torsional bone strength). We used Spearman's rho to identify associations between CV% and time between measurements, child's age or anthropometrics.

Results: After excluding unanalyzable scans (6-10% of scans per bone site), CV%RMS ranged from 4% (total density) to 19% (trabecular content) at the distal radius, 4% (cortical content) to 8% (cortical thickness) at the radius shaft, 2% (total density) to 14% (trabecular content) at the distal tibia and from 2% (cortical content) to 6% (bone strength) at the tibia shaft. Precision errors were within 95% limits-of-agreement across scan quality. Age was associated (rho -0.4 to -0.5, p⟨0.05) with CV% at the tibia.

Conclusion: Bone density outcomes and cortical bone properties appeared most precise (CV%RMS⟨5%) in children.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Bone Density / physiology*
  • Child
  • Cortical Bone / diagnostic imaging*
  • Female
  • Humans
  • Image Interpretation, Computer-Assisted
  • Imaging, Three-Dimensional
  • Male
  • Radius / diagnostic imaging*
  • Tibia / diagnostic imaging*
  • Tomography, X-Ray Computed / methods*

Grants and funding