1-Year Results in Patients Undergoing Transcatheter Aortic Valve Replacement With Failed Surgical Bioprostheses

JACC Cardiovasc Interv. 2017 May 22;10(10):1034-1044. doi: 10.1016/j.jcin.2017.03.018.

Abstract

Objectives: This study evaluated the safety and effectiveness of self-expanding transcatheter aortic valve replacement (TAVR) in patients with surgical valve failure (SVF).

Background: Self-expanding TAVR is superior to medical therapy for patients with severe native aortic valve stenosis at increased surgical risk.

Methods: The CoreValve U.S. Expanded Use Study was a prospective, nonrandomized study that enrolled 233 patients with symptomatic SVF who were deemed unsuitable for reoperation. Patients were treated with self-expanding TAVR and evaluated for 30-day and 1-year outcomes after the procedure. An independent core laboratory was used to evaluate serial echocardiograms for valve hemodynamics and aortic regurgitation.

Results: SVF occurred through stenosis (56.4%), regurgitation (22.0%), or a combination (21.6%). A total of 227 patients underwent attempted TAVR and successful TAVR was achieved in 225 (99.1%) patients. Patients were elderly (76.7 ± 10.8 years), had a Society of Thoracic Surgeons Predicted Risk of Mortality score of 9.0 ± 6.7%, and were severely symptomatic (86.8% New York Heart Association functional class III or IV). The all-cause mortality rate was 2.2% at 30 days and 14.6% at 1 year; major stroke rate was 0.4% at 30 days and 1.8% at 1 year. Moderate aortic regurgitation occurred in 3.5% of patients at 30 days and 7.4% of patients at 1 year, with no severe aortic regurgitation. The rate of new permanent pacemaker implantation was 8.1% at 30 days and 11.0% at 1 year. The mean valve gradient was 17.0 ± 8.8 mm Hg at 30 days and 16.6 ± 8.9 mm Hg at 1 year. Factors significantly associated with higher discharge mean aortic gradients were surgical valve size, stenosis as modality of SVF, and presence of surgical valve prosthesis patient mismatch (all p < 0.001).

Conclusions: Self-expanding TAVR in patients with SVF at increased risk for surgery was associated with a low 1-year mortality and major stroke rate, significantly improved aortic valve hemodynamics, and low rates of moderate and no severe residual aortic regurgitation, with improved quality of life.

Keywords: TAV in SAV; TAVR; prospective trial; surgical valve failure.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve / diagnostic imaging
  • Aortic Valve / physiopathology
  • Aortic Valve / surgery*
  • Aortic Valve Insufficiency / diagnostic imaging
  • Aortic Valve Insufficiency / mortality
  • Aortic Valve Insufficiency / physiopathology
  • Aortic Valve Insufficiency / surgery*
  • Aortic Valve Stenosis / diagnostic imaging
  • Aortic Valve Stenosis / mortality
  • Aortic Valve Stenosis / physiopathology
  • Aortic Valve Stenosis / surgery*
  • Echocardiography
  • Female
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / instrumentation*
  • Heart Valve Prosthesis Implantation / mortality
  • Heart Valve Prosthesis*
  • Hemodynamics
  • Humans
  • Kaplan-Meier Estimate
  • Male
  • Middle Aged
  • Prospective Studies
  • Prosthesis Failure*
  • Quality of Life
  • Recovery of Function
  • Recurrence
  • Reoperation
  • Risk Factors
  • Time Factors
  • Transcatheter Aortic Valve Replacement / adverse effects
  • Transcatheter Aortic Valve Replacement / instrumentation*
  • Transcatheter Aortic Valve Replacement / mortality
  • Treatment Outcome