Flow injection microfluidic device with on-line fluorescent derivatization for the determination of Cr(III) and Cr(VI) in water samples after solid phase extraction

Anal Chim Acta. 2017 Feb 22:955:58-66. doi: 10.1016/j.aca.2016.11.057. Epub 2016 Dec 2.

Abstract

In this paper, a rapid and simple method using magnetic multi-walled carbon nanotubes (MWCNTS), as a solid-phase extraction (SPE) sorbent, was successfully developed for extraction and preconcentration trace amounts of Cr(III) in water samples. The synthesized magnetic-MWCNTs nanocomposite was characterized by scanning electron microscopy (SEM), transmission electron microscopy (TEM), Fourier transform infrared spectroscopy (FT-IR), X-ray photoelectron spectroscopy (XPS) and X-ray diffraction (XRD). A rhodamine derivative (R1) was synthesized and characterized as a highly selective and sensitive fluorescent derivatizing agent for Cr(III). After SPE procedure, Cr(III) analysis was performed by flow injection microfluidic chip with on-line fluorescent derivatization and laser-induced fluorescence (LIF) spectroscopy detection. The parameters, which affected the efficiency of the developed method were investigated and optimized. Under the optimized conditions, the method exhibited a linear dynamic range of 0-10.0 nM, with a detection limit of 0.094 nM and an enrichment factor of 38. Furthermore, real water samples were analyzed and good recoveries were obtained from 91.0 to 101.6%.

Keywords: Flow injection microfluidic chip; Laser-induced fluorescence spectroscopy; Magnetic multi-walled carbon nanotubes; Solid-phase extraction.