Comparison of maximum force to failure of 4 thoracostomy tube connecting devices

Vet Surg. 2017 Feb;46(2):249-254. doi: 10.1111/vsu.12605. Epub 2016 Dec 28.

Abstract

Objective: To compare the maximum force and displacement to failure of 4 different types of thoracostomy tube connecting devices.

Study design: Experimental in vitro study.

Study population: Four types of thoracostomy tube connecting devices (n = 10 each).

Methods: Four different connecting device configurations (10 constructs each) were tested by maximum distraction to failure using a dynamometer: (1) CTTWW-a 3-way connector with a male luer slip attached to a thoracostomy tube by a Christmas tree adapter and secured to the tube with 21 gauge orthopedic wire; (2) CTTWRCW-a 3-way connector with a male luer lock with a rotating collar attached to a tube by a Christmas tree adapter and secured to the tube with 21 gauge orthopedic wire; (3) LVSBC-a Lopez valve attached to a tube with its short-barbed connector; and (4) LVLBC-a Lopez valve attached to a tube with its long-barbed connector.

Results: The maximum distraction force to failure was significantly greater for CTTWRCW (250.9 N; range 143.7-293.6) than CTTWW (132.9 N; range 84.2-224.1), LVLBC (90.8 N; range 74.0-123.4), and LVSBC (54.6 N; range 39.6-164.2). The median displacement to failure of CTTWRCW (150 mm; range 54-190) was significantly longer than that of CTTWW (34.5 mm; range 22-70), LVLBC (32.5 mm; range 24-57), and LVSBC (16 mm; range 11-69).

Conclusion: The CTTWRCW group required greater force to create failure and had a longer displacement to failure, making it a more secure choice for connection to thoracostomy tubes.

Publication types

  • Comparative Study

MeSH terms

  • Animals
  • Biomechanical Phenomena
  • Chest Tubes / veterinary*
  • Equipment Design
  • Pleural Effusion / surgery
  • Pleural Effusion / veterinary
  • Pneumothorax / surgery
  • Pneumothorax / veterinary
  • Thoracostomy / veterinary*