Effects of remote monitoring on clinical outcomes and use of healthcare resources in heart failure patients with biventricular defibrillators: results of the MORE-CARE multicentre randomized controlled trial

Eur J Heart Fail. 2017 Mar;19(3):416-425. doi: 10.1002/ejhf.626. Epub 2016 Aug 28.

Abstract

Aims: The aim of this study was to evaluate the clinical efficacy and safety of remote monitoring in patients with heart failure implanted with a biventricular defibrillator (CRT-D) with advanced diagnostics.

Methods and results: The MORE-CARE trial is an international, prospective, multicentre, randomized controlled trial. Within 8 weeks of de novo implant of a CRT-D, patients were randomized to undergo remote checks alternating with in-office follow-ups (Remote arm) or in-office follow-ups alone (Standard arm). The primary endpoint was a composite of death and cardiovascular (CV) and device-related hospitalization. Use of healthcare resources was also evaluated. A total of 865 eligible patients (mean age 66 ± 10 years) were included in the final analysis (437 in the Remote arm and 428 in the Standard arm) and followed for a median of 24 (interquartile range = 15-26) months. No significant difference was found in the primary endpoint between the Remote and Standard arms [hazard ratio 1.02, 95% confidence interval (CI) 0.80-1.30, P = 0.89] or in the individual components of the primary endpoint (P > 0.05). For the composite endpoint of healthcare resource utilization (i.e. 2-year rates of CV hospitalizations, CV emergency department admissions, and CV in-office follow-ups), a significant 38% reduction was found in the Remote vs. Standard arm (incidence rate ratio 0.62, 95% CI 0.58-0.66, P < 0.001) mainly driven by a reduction of in-office visits.

Conclusions: In heart failure patients implanted with a CRT-D, remote monitoring did not reduce mortality or risk of CV or device-related hospitalization. Use of healthcare resources was significantly reduced as a result of a marked reduction of in-office visits without compromising patient safety.

Trial registration: NCT00885677.

Keywords: Cardiac resynchronization therapy; Costs; Healthcare; Heart failure; Outcome; Remote monitoring.

Publication types

  • Multicenter Study
  • Randomized Controlled Trial

MeSH terms

  • Aftercare / methods*
  • Aged
  • Cardiac Resynchronization Therapy Devices*
  • Cardiac Resynchronization Therapy*
  • Cardiovascular Diseases
  • Emergency Service, Hospital / statistics & numerical data*
  • Female
  • Heart Failure / therapy*
  • Hospitalization / statistics & numerical data*
  • Humans
  • Male
  • Middle Aged
  • Monitoring, Ambulatory*
  • Mortality*
  • Prospective Studies
  • Telemedicine / methods*

Associated data

  • ClinicalTrials.gov/NCT00885677