A Comparison of Mindray BC-6800, Sysmex XN-2000, and Beckman Coulter LH750 Automated Hematology Analyzers: A Pediatric Study

J Clin Lab Anal. 2016 Nov;30(6):1128-1134. doi: 10.1002/jcla.21992. Epub 2016 May 17.

Abstract

Background: Modern automated laboratory hematology analyzers allow the measurement of over 30 different hematological parameters useful in the diagnostic and clinical interpretation of patient symptoms. They use different methods to measure the same parameters. Thus, a comparison of complete blood count made by Mindray BC-6800, Sysmex XN-2000 and Beckman Coulter LH750 was performed.

Materials and methods: A comparison of results obtained by automated analysis of 807 anticoagulated blood samples from children and 125 manual microscopic differentiations were performed. This comparative study included white blood cell count, red blood cell count, and erythrocyte indices, as well as platelet count.

Results: The present study showed a poor level of agreement between white blood cell enumeration and differentiation of the three automated hematology analyzers under comparison. A very good agreement was found when comparing manual blood smear and automated granulocytes, monocytes, and lymphocytes differentiation. Red blood cell evaluation showed better agreement than white blood cells between the studied analyzers.

Conclusion: To conclude, studied instruments did not ensure satisfactory interchangeability and did not facilitate a substitution of one analyzer by another.

Keywords: automated hematology analyzer; comparison; complete blood count; pediatric hematology.

MeSH terms

  • Adolescent
  • Automation, Laboratory*
  • Blood Cell Count / instrumentation*
  • Blood Cell Count / methods*
  • Blood Cells / cytology
  • Blood Cells / physiology*
  • Child
  • Child, Preschool
  • Erythrocyte Count
  • Female
  • Hematology / instrumentation*
  • Hematology / methods*
  • Humans
  • Infant
  • Leukocyte Count
  • Male
  • Platelet Count
  • Regression Analysis
  • Reproducibility of Results