Initial characterization of a gel patch dosimeter for in vivo dosimetry

Phys Med Biol. 2016 May 21;61(10):N240-8. doi: 10.1088/0031-9155/61/10/N240. Epub 2016 Apr 18.

Abstract

In vivo dosimetry is a greatly underutilized tool for patient safety in clinical external beam radiotherapy treatments, despite being recommended by several national and international organizations (AAPM, ICRU, IAEA, NACP). The reasons for this underutilization mostly relate to the feasibility and cost of in vivo dosimetry methods. Due to the increase in the number of beam angles and dose per fraction in modern treatments, there is a compelling need for a novel dosimeter that is robust and affordable while able to operate properly in these complex conditions. This work presents a gel patch dosimeter as a novel method of in vivo dosimetry. DEFGEL, a 6% T normoxic polyacrylamide gel, was injected into 1 cm thick acrylic molds to create 1 cm thick small cylindrical patch dosimeters. To evaluate the change in optical density due to radiation induced polymerization, dosimeters were scanned before and after irradiation using an in-house developed laser densitometer. The dose-responses of three separate batches of gel were evaluated and compared to check for linearity and repeatability. The response development time was evaluated to ensure that the patch dosimeter could be high throughput. Additionally, the potential of this system to be used as an in vivo dosimeter was tested with a clinically relevant end-to-end in vivo phantom test. All irradiations were performed with a Varian Clinac 21EX at the University of Wisconsin Medical Radiation Research Center (UWMRRC). The dose-response of all three batches of gel was found to be linear within the range of 2-20 Gy. At doses below 0.5 Gy the statistical uncertainties were prohibitively large to make quantitative assessments of the results. The three batches demonstrated good repeatability in the range of 2 Gy to up to 10 Gy, with only slight variations in response at higher doses. For low doses the dosimeter fully developed within an hour while at higher doses they fully developed within four hours. During the in vivo phantom test the predicted patch absorbed dose was 4.23 Gy while the readout dose was evaluated to be 4.37 Gy, which corresponds to a 3.2% discrepancy. The dosimeter and densitometer pairing shows promise as an in vivo dosimetry system, especially for hypofractionated or MRI-guided radiotherapy treatments where higher doses are prescribed.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Acrylic Resins / radiation effects
  • Gels / radiation effects
  • Humans
  • In Vivo Dosimetry / methods*
  • Phantoms, Imaging
  • Radiation Dosimeters*
  • Radiotherapy Dosage

Substances

  • Acrylic Resins
  • Gels
  • polyacrylamide