Reliability of Single-Use PEEP-Valves Attached to Self-Inflating Bags during Manual Ventilation of Neonates--An In Vitro Study

PLoS One. 2016 Feb 25;11(2):e0150224. doi: 10.1371/journal.pone.0150224. eCollection 2016.

Abstract

Introduction: International resuscitation guidelines suggest to use positive end-expiratory pressure (PEEP) during manual ventilation of neonates. Aim of our study was to test the reliability of self-inflating bags (SIB) with single-use PEEP valves regarding PEEP delivery and the effect of different peak inflation pressures (PIP) and ventilation rates (VR) on the delivered PEEP.

Methods: Ten new single-use PEEP valves from 5 manufacturers were tested by ventilating an intubated 1 kg neonatal manikin containing a lung model with a SIB that was actuated by an electromechanical plunger device. Standard settings: PIP 20 cmH2O, VR 60/min, flow 8 L/min. PEEP settings of 5 and 10 cmH2O were studied. A second test was conducted with settings of PIP 40 cmH2O and VR 40/min. The delivered PEEP was measured by a respiratory function monitor (CO2SMO+).

Results: Valves from one manufacturer delivered no relevant PEEP and were excluded. The remaining valves showed a continuous decay of the delivered pressure during expiration. The median (25th and 75th percentile) delivered PEEP with standard settings was 3.4(2.7-3.8) cmH2O when set to 5 cmH2O and 6.1(4.9-7.1) cmH2O when set to 10 cmH2O. Increasing the PIP from 20 to 40 cmH2O led to a median (25th and 75th percentile) decrease in PEEP to 2.3(1.8-2.7) cmH2O and 4.3(3.2-4.8) cmH2O; changing VR from 60 to 40/min led to a PEEP decrease to 2.8(2.1-3.3) cmH2O and 5.0(3.5-6.2) cmH2O for both PEEP settings.

Conclusion: Single-use PEEP valves do not reliably deliver the set PEEP. PIP and VR have an effect on the delivered PEEP. Operators should be aware of these limitations when manually ventilating neonates.

Publication types

  • Evaluation Study

MeSH terms

  • Cardiopulmonary Resuscitation / instrumentation*
  • Cardiopulmonary Resuscitation / methods*
  • Equipment Design
  • Humans
  • Infant, Newborn
  • Lung / physiopathology
  • Manikins
  • Monitoring, Physiologic / methods
  • Positive-Pressure Respiration / instrumentation*
  • Positive-Pressure Respiration / methods
  • Reproducibility of Results

Grants and funding

The authors have no support or funding to report.