Orbital color Doppler ultrasound as noninvasive tool in the diagnosis of anterior-draining carotid-cavernous fistula

Radiol Med. 2016 Apr;121(4):301-7. doi: 10.1007/s11547-015-0607-0. Epub 2015 Dec 6.

Abstract

Purpose: To investigate the role of orbital color Doppler ultrasound (OCDUS) in the diagnosis of carotid-cavernous fistula (CCF) with anterior drainage and particularly whether a negative OCDUS could avoid an invasive diagnostic cerebral angiography (DSA).

Materials and methods: Twenty-two consecutive patients with ophthalmic signs suspecting CCF were submitted to ophthalmologic examination, OCDUS and DSA. CCF diagnosis with OCDUS was based on the finding of a reversed, arterialized and low-resistive-index (RI <0.5) blood flow in the superior ophthalmic vein (SOV). Sensibility, specificity, PPV, NPV, and accuracy of OCDUS were calculated considering both patients and eyes, using DSA as gold standard.

Results: DSA demonstrated 20 CCFs in 18 patients. Considering the patients, in 18/22 CCF diagnosis was positive at OCDUS and DSA while 4/22 were negative at both. Considering the eyes, in 24/43 CCF diagnosis was positive at both DSA and OCDUS (total eyes = 43, due to one case of SOV thrombosis). In 19/43 eyes diagnosis was negative at both OCDUS and DSA. So sensitivity, specificity, PPV, NPV, and accuracy of OCDUS in the patients and eyes analysis were all 100 %.

Conclusions: OCDUS is a reliable, noninvasive tool in the diagnosis of CCF; a negative OCDUS could avoid an invasive DSA in patients suspected for anterior-draining CCF.

Keywords: Carotid-cavernous sinus fistula; Cerebral angiography; Color Doppler ultrasonography; Eye diseases; Ultrasonography.

MeSH terms

  • Carotid-Cavernous Sinus Fistula / diagnostic imaging*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Retrospective Studies
  • Ultrasonography, Doppler, Color*