Transcatheter Aortic and Mitral Valve-in-Valve Implantation for Failed Surgical Bioprosthetic Valves: An 8-Year Single-Center Experience

JACC Cardiovasc Interv. 2015 Nov;8(13):1735-44. doi: 10.1016/j.jcin.2015.08.012. Epub 2015 Oct 14.

Abstract

Objectives: We report our 8-year experience in transcatheter aortic and mitral valve-in-valve (VinV) implantation.

Background: Feasibility and good early outcomes associated with transcatheter aortic and mitral VinV implantation into failed surgical bioprostheses have been confirmed, but the mid-term and long-term outcomes of transcatheter aortic and mitral VinV is unknown.

Methods: A total of 73 patients with aortic (n = 42) and mitral (n = 31) bioprosthetic valve dysfunction underwent transcatheter VinV implantation between April 2007 and December 2013. Edwards balloon-expandable transcatheter valves (Edwards Lifesciences Inc., Irvine, California) were used. Median follow-up was 2.52 years with a maximum of 8 years.

Results: Seventy-two patients (mean age 79.7 ± 9.4 years, 32 women) underwent successful VinV implantation (success rate 98.6%). At 30 days, all-cause mortality was 1.4%, disabling stroke 1.4%, life-threatening bleeding 4.1%, acute kidney injury requiring hemodialysis 2.7%, and coronary artery obstruction requiring intervention 1.4%. No patient had greater than mild paravalvular leak. Estimated survival rates were 88.9%, 79.5%, 69.8%, 61.9%, and 40.5% at 1, 2, 3, 4, and 5 years, respectively. The small surgical valve size (19 and 21 mm) was an independent risk factor for reduced survival in aortic VinV patients. At 2-year follow-up, 82.8% of aortic and 100% of mitral VinV patients were in New York Heart Association functional class I or II.

Conclusions: Transcatheter VinV for failed surgical bioprostheses can be performed safely with a high success rate and minimal early mortality and morbidity. Transcatheter VinV provides encouraging mid-term clinical outcomes in this high-risk elderly cohort of patients. Transcatheter VinV is an acceptable alternative therapy for failed aortic or mitral bioprostheses in selected high-risk patients.

Keywords: aortic valve; bioprosthesis; mitral valve; transcatheter; valve-in-valve.

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve / physiopathology
  • Aortic Valve / surgery*
  • Bioprosthesis*
  • British Columbia
  • Cardiac Catheterization* / adverse effects
  • Cardiac Catheterization* / instrumentation
  • Cardiac Catheterization* / methods
  • Cardiac Catheterization* / mortality
  • Feasibility Studies
  • Female
  • Heart Valve Diseases / diagnosis
  • Heart Valve Diseases / mortality
  • Heart Valve Diseases / physiopathology
  • Heart Valve Diseases / surgery
  • Heart Valve Diseases / therapy*
  • Heart Valve Prosthesis Implantation* / adverse effects
  • Heart Valve Prosthesis Implantation* / instrumentation
  • Heart Valve Prosthesis Implantation* / methods
  • Heart Valve Prosthesis Implantation* / mortality
  • Heart Valve Prosthesis*
  • Humans
  • Kaplan-Meier Estimate
  • Male
  • Mitral Valve / physiopathology
  • Mitral Valve / surgery*
  • Patient Selection
  • Prosthesis Design
  • Prosthesis Failure*
  • Recovery of Function
  • Retreatment
  • Risk Factors
  • Time Factors
  • Treatment Failure