Usefulness of IDEAL T2 imaging for homogeneous fat suppression and reducing susceptibility artefacts in brachial plexus MRI at 3.0 T

Radiol Med. 2016 Jan;121(1):45-53. doi: 10.1007/s11547-015-0576-3. Epub 2015 Aug 8.

Abstract

Purpose: To quantitatively and qualitatively compare fat-suppressed MR imaging quality using iterative decomposition of water and fat with echo asymmetry and least-squares estimation (IDEAL) with that using frequency-selective fat-suppressed (FSFS) T2 images of the brachial plexus at 3.0 T.

Materials and methods: Prospective MR image analysis was performed in 40 volunteers and 40 patients at a single centre. Oblique-sagittal and coronal IDEAL fat-suppressed T2 images and FSFS T2 images were compared. Visual assessment was performed by two independent musculoskeletal radiologists with respect to: (1) susceptibility artefacts around the neck, (2) homogeneity of fat suppression, (3) image sharpness and (4) tissue resolution contrast of pathologies. The signal-to-noise ratios (SNR) for each image sequence were assessed.

Results: Compared to FSFS sequences, IDEAL fat-suppressed T2 images significantly reduced artefacts around the brachial plexus and significantly improved homogeneous fat suppression (p < 0.05). IDEAL significantly improved sharpness and lesion-to-tissue contrast (p < 0.05). The mean SNRs were significantly improved on T2-weighted IDEAL images (p < 0.05).

Conclusion: IDEAL technique improved image quality by reducing artefacts around the brachial plexus while maintaining a high SNR and provided superior homogeneous fat suppression than FSFS sequences.

Keywords: Artefact; Brachial plexus; Fat saturation; IDEAL; MRI.

MeSH terms

  • Adipose Tissue / anatomy & histology
  • Adolescent
  • Adult
  • Aged
  • Artifacts
  • Brachial Plexus Neuropathies / diagnosis*
  • Case-Control Studies
  • Humans
  • Image Enhancement / methods*
  • Image Processing, Computer-Assisted / methods*
  • Magnetic Resonance Imaging / methods*
  • Middle Aged
  • Signal-To-Noise Ratio