Importance: Risk stratification and management paradigms for patients with cardiovascular implantable electronic devices (CIEDs) requiring radiotherapy (RT) vary widely and are based on limited clinical data.
Objective: To identify the incidence and predictors of CIED malfunction and describe associated clinical consequences in a large cohort of patients treated with photon- and electron-based RT.
Design, setting, and participants: Retrospective analysis of all patients with a functioning CIED who underwent RT between August 2005 and January 2014 with CIED interrogation data following RT at an academic cancer center. We identified 249 courses of photon- and electron-based RT in 215 patients (123 pacemakers [57%]; 92 implantable cardioverter-defibrillators [43%]). Substantial neutron production was generated in 71 courses (29%).
Exposure: Implantation of CIED with subsequent therapeutic radiation treatment (neutron producing with 15- or 18-MV photons and non-neutron producing with electrons, GammaKnife, or 6-MV photons).
Main outcomes and measures: Malfunction of CIED, characterized as single-event upset (data loss, parameter resets, unrecoverable resets), and delayed effects including signal interference, pacing threshold changes, and premature battery depletion.
Results: Malfunction of CIED attributable to RT occurred during 18 courses (7%), with 15 CIEDs experiencing single-event upsets, and 3, transient signal interference. All single-event upsets occurred during neutron-producing RT, at a rate of 21%, 10%, and 34% per neutron-producing course for CIEDs, pacemakers, and implantable cardioverter-defibrillators, respectively. No single-event upsets were found among 178 courses of non-neutron-producing RT. Incident CIED dose did not correlate with device malfunction. Patients treated to the abdomen and pelvis region were more likely to undergo a single-event upset (hazard ratio, 5.2 [95% CI, 1.2-22.6]; P = .03). Six patients with a CIED parameter reset developed clinical symptoms: 3 experienced hypotension and/or bradycardia, 2 experienced abnormal chest ticking consistent with pacemaker syndrome, and 1 developed congestive heart failure. The 3 episodes of signal interference did not result in clinical effects. No delayed malfunctions were directly attributed to RT.
Conclusions and relevance: In a cohort of contemporary CIEDs, all cases of single-event upset malfunction occurred in the setting of notable neutron production, at a rate of 21% for neutron-producing RT and 0% for non-neutron-producing RT. Where clinically feasible, the use of non-neutron-producing RT is recommended. Given the lack of correlation between CIED malfunction and incident dose observed up to 5.4 Gy, invasive CIED relocation procedures in these settings can be minimized.