Evaluation of HER2 (ERBB2) gene amplification or protein expression is standard of care in breast (BR) and advanced stage gastro-oesophageal cancers to identify patients eligible for anti-HER2 therapies. Here, we evaluate a rapid fluorescence in situ hybridisation (FISH) technology (HER2 instant quality (IQ) FISH pharmDx Kit) for detection of HER2 in patients with BR and gastro-oesophageal cancer using 30 FFPE samples that had been previously evaluated with the PathVysion HER2 DNA Probe Kit. Cases were scored as positive (HER2:CEN-17 ≥2.0), negative (HER2:CEN-17 <2.0) or equivocal according to the ASCO/CAP 2013 BR cancer guidelines. Ten samples were positive for HER2 amplification while 20 were negative; none were equivocal. The IQ FISH was able to detect low level amplification (HER2:CEN-17 ratio 2.4). The HER2 IQ FISH pharmDx Kit is a FDA approved kit that offers a rapid turnaround time (approximately 3.5 h) and in our laboratory was 100% concordant with prior PathVysion results.
Keywords: BREAST CANCER; FISH; MOLECULAR PATHOLOGY.
Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://group.bmj.com/group/rights-licensing/permissions.