Simultaneous whole-body time-of-flight 18F-FDG PET/MRI: a pilot study comparing SUVmax with PET/CT and assessment of MR image quality

Clin Nucl Med. 2015 Jan;40(1):1-8. doi: 10.1097/RLU.0000000000000611.

Abstract

Purpose: The recent introduction of hybrid PET/MRI scanners in clinical practice has shown promising initial results for several clinical scenarios. However, the first generation of combined PET/MRI lacks time-of-flight (TOF) technology. Here we report the results of the first patients to be scanned on a completely novel fully integrated PET/MRI scanner with TOF.

Materials and methods: We analyzed data from patients who underwent a clinically indicated F FDG PET/CT, followed by PET/MRI. Maximum standardized uptake values (SUVmax) were measured from F FDG PET/MRI and F FDG PET/CT for lesions, cerebellum, salivary glands, lungs, aortic arch, liver, spleen, skeletal muscle, and fat. Two experienced radiologists independently reviewed the MR data for image quality.

Results: Thirty-six patients (19 men, 17 women, mean [±standard deviation] age of 61 ± 14 years [range: 27-86 years]) with a total of 69 discrete lesions met the inclusion criteria. PET/CT images were acquired at a mean (±standard deviation) of 74 ± 14 minutes (range: 49-100 minutes) after injection of 10 ± 1 mCi (range: 8-12 mCi) of F FDG. PET/MRI scans started at 161 ± 29 minutes (range: 117 - 286 minutes) after the F FDG injection. All lesions identified on PET from PET/CT were also seen on PET from PET/MRI. The mean SUVmax values were higher from PET/MRI than PET/CT for all lesions. No degradation of MR image quality was observed.

Conclusion: The data obtained so far using this investigational PET/MR system have shown that the TOF PET system is capable of excellent performance during simultaneous PET/MR with routine pulse sequences. MR imaging was not compromised. Comparison of the PET images from PET/CT and PET/MRI show no loss of image quality for the latter. These results support further investigation of this novel fully integrated TOF PET/MRI instrument.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Female
  • Fluorodeoxyglucose F18 / administration & dosage
  • Fluorodeoxyglucose F18 / pharmacokinetics*
  • Humans
  • Magnetic Resonance Imaging*
  • Male
  • Middle Aged
  • Multimodal Imaging*
  • Pilot Projects
  • Positron-Emission Tomography*
  • Radiopharmaceuticals / administration & dosage
  • Radiopharmaceuticals / pharmacokinetics*
  • Reproducibility of Results
  • Tomography, X-Ray Computed*
  • Whole Body Imaging*