Comparison of COSMED'S FitMate™ and K4b2 metabolic systems reliability during graded cycling exercise

Scand J Clin Lab Invest. 2014 Nov;74(8):722-4. doi: 10.3109/00365513.2014.930711.

Abstract

This study compared the reliability of the Cosmed FitMate™ and K4b2 metabolic systems during light to heavy steady state exercise. Expired gas, ventilation were recorded in 50 subjects, using in a random order among four sessions, either the FitMate™ or the Cosmed K4b2. No differences in oxygen consumption were observed between the two systems whatever the intensity. Intraclass correlation were high for both analyzers (respectively for the FitMate™ system and the Cosmed K4b2; ICC: 0.76-0.88 vs. 0.88-0.95). The FitMate™ metabolic system could be a useful reliable and easy-to-use metabolic system in energy expenditure measurement.

Keywords: Methodology; energy expenditure; exercise test; oxygen consumption; reliability.