Four-dimensional computed tomography based respiratory-gated radiotherapy with respiratory guidance system: analysis of respiratory signals and dosimetric comparison

Biomed Res Int. 2014:2014:306021. doi: 10.1155/2014/306021. Epub 2014 Sep 7.

Abstract

Purpose: To investigate the effectiveness of respiratory guidance system in 4-dimensional computed tomography (4 DCT) based respiratory-gated radiation therapy (RGRT) by comparing respiratory signals and dosimetric analysis of treatment plans.

Methods: The respiratory amplitude and period of the free, the audio device-guided, and the complex system-guided breathing were evaluated in eleven patients with lung or liver cancers. The dosimetric parameters were assessed by comparing free breathing CT plan and 4 DCT-based 30-70% maximal intensity projection (MIP) plan.

Results: The use of complex system-guided breathing showed significantly less variation in respiratory amplitude and period compared to the free or audio-guided breathing regarding the root mean square errors (RMSE) of full inspiration (P = 0.031), full expiration (P = 0.007), and period (P = 0.007). The dosimetric parameters including V(5 Gy), V(10 Gy), V(20 Gy), V(30 Gy), V(40 Gy), and V(50 Gy) of normal liver or lung in 4 DCT MIP plan were superior over free breathing CT plan.

Conclusions: The reproducibility and regularity of respiratory amplitude and period were significantly improved with the complex system-guided breathing compared to the free or the audio-guided breathing. In addition, the treatment plan based on the 4D CT-based MIP images acquired with the complex system guided breathing showed better normal tissue sparing than that on the free breathing CT.

Publication types

  • Comparative Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Dose-Response Relationship, Radiation
  • Female
  • Four-Dimensional Computed Tomography / methods*
  • Humans
  • Male
  • Middle Aged
  • Radiotherapy Dosage
  • Radiotherapy*
  • Respiration
  • Respiratory-Gated Imaging Techniques / methods*
  • Signal Processing, Computer-Assisted*