The concept of a monolithic Ti-Ni spinal rod with variable flexural stiffness is proposed to reduce the risks associated with spinal fusion. The variable stiffness is conferred to the rod using the Joule-heating local annealing technique. The annealing temperature and the mechanical properties' distributions resulted from this thermal treatment are numerically modeled and experimentally measured. To illustrate the possible applications of such a modeling approach, two case studies are presented: (a) optimization of the Joule-heating strategy to reduce annealing time, and (b) modulation of the rod's overall flexural stiffness using partial annealing. A numerical model of a human spine coupled with the model of the variable flexural stiffness spinal rod developed in this work can ultimately be used to maximize the stabilization capability of spinal instrumentation, while simultaneously decreasing the risks associated with spinal fusion.
Keywords: Finite element analysis; Instrumentation with variable flexural stiffness; Local Joule-heating annealing; Shape memory alloy; Spinal fusion.
Copyright © 2014 IPEM. Published by Elsevier Ltd. All rights reserved.