Effect of increasing the sampling interval to 2 seconds on the radiation dose and accuracy of CT perfusion of the head and neck

J Comput Assist Tomogr. 2014 May-Jun;38(3):469-73. doi: 10.1097/RCT.0000000000000066.

Abstract

Purpose: To evaluate the effect of increasing the sampling interval from 1 second (1 image per second) to 2 seconds (1 image every 2 seconds) on computed tomographic (CT) perfusion (CTP) of head and neck tumors.

Materials and methods: Twenty patients underwent CTP studies of head and neck tumors with images acquired in cine mode for 50 seconds using sampling interval of 1 second. Using deconvolution-based software, analysis of CTP was done with sampling interval of 1 second and then 2 seconds. Perfusion maps representing blood flow, blood volume, mean transit time, and permeability surface area product (PS) were obtained. Quantitative tumor CTP values were compared between the 2 sampling intervals. Two blinded radiologists compared the subjective quality of CTP maps using a 3-point scale between the 2 sampling intervals. Radiation dose parameters were recorded for the 2 sampling interval rates.

Results: No significant differences were observed between the means of the 4 perfusion parameters generated using both sampling intervals; all P >0.05. The 95% limits of agreement between the 2 sampling intervals were -65.9 to 48.1) mL/min per 100 g for blood flow, -3.6 to 3.1 mL/100 g for blood volume, -2.9 to 3.8 seconds for mean transit time, and -10.0 to 12.5 mL/min per 100 g for PS. There was no significant difference between the subjective quality scores of CTP maps obtained using the 2 sampling intervals; all P > 0.05. Radiation dose was halved when sampling interval increased from 1 to 2 seconds.

Conclusions: Increasing the sampling interval rate to 1 image every 2 seconds does not compromise the image quality and has no significant effect on quantitative perfusion parameters of head and neck tumors. The radiation dose is halved.

MeSH terms

  • Head and Neck Neoplasms / diagnostic imaging*
  • Humans
  • Perfusion Imaging / methods*
  • Radiation Dosage*
  • Radiation Protection / methods*
  • Radiographic Image Interpretation, Computer-Assisted / methods*
  • Reproducibility of Results
  • Sample Size
  • Sensitivity and Specificity
  • Tomography, X-Ray Computed / methods*