His or para-His pacing preserves left ventricular function in atrioventricular block: a double-blind, randomized, crossover study

Europace. 2014 Aug;16(8):1189-96. doi: 10.1093/europace/euu011. Epub 2014 Feb 7.

Abstract

Aims: To compare left ventricular function after a long-term His or para-His pacing (HP) and right ventricular septal pacing (RVSP) in patients with atrioventricular block (AVB).

Methods and results: We included consecutive patients with AVB, a narrow QRS < 120 ms, and a preserved left ventricular ejection fraction (LVEF) >0.40, in a prospective, randomized, double-blinded, crossover design. All patients were treated with 12 months HP and 12 months RVSP. A total of 38 patients [mean age, 67 ± 10 years; 30 (79%) men] were included. The primary endpoint was LVEF, which was significantly lower after a 12 months RVSP (0.50 ± 0.11) than after 12 months of HP (0.55 ± 0.10), P = 0.005. We measured the difference in time-to-peak systolic velocity between opposite basal segments in the apical views by using tissue Doppler imaging. In the four-chamber view, the difference was 58 (±7) ms after RVSP and 49 (±7) ms after HP, P = 0.27; in the two-chamber view, the difference was 45 (±5) ms after RVSP and 31 ±(4) ms after HP, P = 0.02, and in the apical long-axis view, the difference was 63 (±6) after RVSP and 44 (±7) after HP, P = 0.03. There was no difference in New York Heart Association class, 6-min hall walk test, quality-of-life assessments, or device-related complications. The mean threshold was significantly higher in HP leads than in RVSP leads.

Conclusion: His or para-His pacing preserves LVEF and mechanical synchrony as compared with RVSP after 12 months pacing in patients with AVB, narrow QRS, and LVEF > 0.40.

Trial registration: ClinicalTrials.gov NCT01019213.

Keywords: Atrioventricular block; His-pacing; Left ventricular function; Pacing.

Publication types

  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Atrioventricular Block / diagnosis
  • Atrioventricular Block / physiopathology
  • Atrioventricular Block / therapy*
  • Bundle of His / diagnostic imaging
  • Bundle of His / physiopathology*
  • Cardiac Resynchronization Therapy / adverse effects
  • Cardiac Resynchronization Therapy / methods*
  • Cardiac Resynchronization Therapy Devices
  • Cross-Over Studies
  • Denmark
  • Double-Blind Method
  • Echocardiography, Doppler
  • Equipment Design
  • Equipment Failure
  • Exercise Test
  • Female
  • Humans
  • Male
  • Middle Aged
  • Prospective Studies
  • Quality of Life
  • Stroke Volume
  • Time Factors
  • Treatment Outcome
  • Ventricular Function, Left*
  • Ventricular Function, Right
  • Ventricular Septum / physiopathology

Associated data

  • ClinicalTrials.gov/NCT01019213