New phase-based B1 mapping method using two-dimensional spin-echo imaging with hyperbolic secant pulses

Magn Reson Med. 2015 Jan;73(1):170-81. doi: 10.1002/mrm.25110. Epub 2014 Jan 23.

Abstract

Purpose: To propose a new phase-based B1-mapping method that exploits phase information created by hyperbolic secant (HS) pulses in conventional 2D spin-echo imaging.

Methods: In this B1-mapping method, HS pulses are used to accomplish π/2 excitation and π refocusing in standard multislice spin-echo imaging. When setting the ratio of pulse lengths of the π/2 and π HS pulses to 2:1, the spin-echo phase is independent of offset frequency and varies as a function of B1 strength. To eliminate undesired phase accumulations induced by unknown factors other than the B1 strength, two spin-echo images are acquired using HS pulses applied with opposite frequency-sweep directions, and the resulting phase images are subtracted from each other. To demonstrate the performance of the proposed method, phantom and in vivo experiments were performed using a surface coil and a volume coil.

Results: The B1 maps obtained by using the proposed method were in accordance with the B1 maps obtained using previous methods in both phantom and in vivo experiments.

Conclusion: The proposed method is easy to implement without any sequence modification, is insensitive to B0 inhomogeneity and chemical shift, and is robust in a reasonably wide range of B1 field strength.

Keywords: B1 mapping; RF field mapping; flip-angle mapping; frequency-modulation pulse; hyperbolic secant pulse; spin-echo imaging.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms*
  • Brain / anatomy & histology*
  • Echo-Planar Imaging / methods*
  • Humans
  • Image Enhancement / methods*
  • Image Interpretation, Computer-Assisted / methods*
  • Numerical Analysis, Computer-Assisted
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Signal Processing, Computer-Assisted*
  • Spin Labels

Substances

  • Spin Labels