Use of LC-MS for the quality control of radiopharmaceuticals: example of [(18)F]ML10

J Labelled Comp Radiopharm. 2013 May 30;56(6):330-3. doi: 10.1002/jlcr.3023. Epub 2013 Feb 21.

Abstract

[(18)F]ML10 is a promising novel low molecular weight positron emission tomography probe for apoptosis. As part of the quality control to support clinical studies for cancer therapy monitoring in the GSK Clinical Imaging Centre, a simple and sensitive liquid chromatography mass spectrometry method has been developed and validated for the quantification of total ML10 and impurity content in the final product. Chromatographic separation of ML10 and its radiolabelling precursor and impurities was achieved. Mass curves were constructed from a concentration range of ML10 and known impurities and were linear. Quantification was achieved by comparison of the area under the curve for ML10 content (m/z = 205) and the mass curve. The method was validated over a concentration range of 0.1-1 µg/ml.

Keywords: LCMS; PET; apoptosis; quality control.

MeSH terms

  • Fluorine Radioisotopes / chemistry
  • Fluorine Radioisotopes / standards*
  • Gas Chromatography-Mass Spectrometry / methods*
  • Methylmalonic Acid / analogs & derivatives*
  • Methylmalonic Acid / chemical synthesis
  • Methylmalonic Acid / chemistry
  • Quality Control*
  • Radiopharmaceuticals / chemistry
  • Radiopharmaceuticals / standards*

Substances

  • 5-fluoropentyl-2-methylmalonic acid
  • Fluorine Radioisotopes
  • Radiopharmaceuticals
  • Methylmalonic Acid