Two-pass dual-energy CT imaging for simultaneous detection, characterization, and volume measurement of urinary stones with excretory-phase CT urography alone: a phantom study

Jpn J Radiol. 2013 Jun;31(6):393-400. doi: 10.1007/s11604-013-0209-5. Epub 2013 May 4.

Abstract

Purpose: To evaluate if two-pass dual-energy CT imaging--i.e., simultaneous three-material and two-material decomposition analysis--can depict and characterize urinary stones in various concentrations of iodine solution in vitro.

Materials and methods: Twelve urinary stones were scanned with a dual-source CT scanner. First, each stone (in a saline-filled tube) underwent single- and dual-energy mode CT scans in order to measure the volume of the stone. Each stone was then placed in various concentrations of contrast medium and scanned in dual-energy mode to calculate its volume via three-material decomposition analysis. Two-pass dual-energy CT imaging analysis software for the Matlab environment, which was developed specifically to process simultaneous three-material and two-material decomposition, was applied to characterize and calculate the volume of each stone.

Results: Although the virtual non-contrast images from three-material decomposition analysis clearly visualized all of the stones in contrast medium with up to 80 mgI/mL, the volumes of the uric acid stones were overestimated. Two-pass dual-energy CT imaging was able to depict and characterize non-uric-acid stones in diluted contrast medium with up to 80 mgI/mL, whereas uric acid stones were correctly evaluated in diluted contrast medium with 40 mgI/mL or less.

Conclusions: Two-pass dual-energy CT imaging is able to depict and characterize urinary stones in contrast medium.

MeSH terms

  • Contrast Media
  • Humans
  • Phantoms, Imaging*
  • Predictive Value of Tests
  • Sensitivity and Specificity
  • Tomography, X-Ray Computed / methods*
  • Urinary Calculi / diagnostic imaging*
  • Urography*