Objectives: To assess the usefulness of the MGuard stent in patients with ST-segment elevation myocardial infarction (STEMI) in whom a high thrombus burden persists after manual aspiration.
Background: In some patients with STEMI, a high thrombus burden may persist after manual aspiration. These patients may be at high risk of distal embolization and therefore impaired myocardial reperfusion. The MGuard is a novel mesh-covered stent designed to minimize thrombus embolization.
Methods: Single-arm, prospective registry of patients with STEMI and high thrombus burden after aggressive thrombus aspiration treated with the MGuard stent. High thrombus burden was defined as thrombus burden grade 4 or 5 according to the TIMI score. Lesions with a side branch ≥2 mm and patients with cardiogenic shock were not included. The study end-points were proportion of final TIMI 3 flow, normal myocardial blush, and complete ST-segment resolution.
Results: Fifty-six patients were included. After MGuard stent implantation >85% of cases had thrombus score = 0. Final TIMI 3 flow was achieved in 82% of cases, normal myocardial blush in 55%, and complete ST-segment resolution in 59%. Occlusion of a side branch (<2 mm) occurred in 2 cases (3.5%), embolization to a distal branch in 5 cases (8.9%), and transient no-reflow in 4 cases (7.1%). Major adverse cardiac events rate at 9 months was 3.6%, including 1 definite acute stent thrombosis and 1 target-vessel revascularization.
Conclusions: The MGuard stent may be useful to prevent distal embolization in patients with STEMI and high thrombus burden despite mechanical aspiration.