Comparison of standard forward-viewing mode versus ultrawide-viewing mode of a novel colonoscopy platform: a prospective, multicenter study in the detection of simulated polyps in an in vitro colon model (with video)

Gastrointest Endosc. 2013 Mar;77(3):472-9. doi: 10.1016/j.gie.2012.12.011.

Abstract

Background: Although colonoscopy is the criterion standard for detecting colorectal adenomas and cancers, a significant percentage of adenomas are missed.

Objective: To compare forward-viewing with ultrawide-viewing colonoscopy in the detection of simulated colon polyps in an in vitro colon model.

Design: Prospective, multicenter.

Setting: Six endoscopy units (3 in the United States and 3 in Israel).

Patients: In vitro colon model with simulated colon polyps (n = 21 metallic beads).

Interventions: Detection of simulated colon polyps on colonoscope withdrawal.

Main outcome measurements: Incremental detection of simulated colon polyps and endoscopist evaluation of the usability, visibility, and maneuverability of ultrawide-viewing colonoscopy.

Results: On forward-viewing colonoscopy, the number of simulated polyps (mean ± standard deviation) detected per endoscopist was 11.1 ± 2.3 polyps, a 52.9% detection rate. Simulated polyp detection rates per colon segment were 3.0 ± 0.93 (60.0%) right colon, 2.4 ± 0.87 (48.0%) transverse colon, and 5.7 ± 1.5 (51.8%) left colon. On ultrawide-viewing colonoscopy, the simulated polyp detection rate per endoscopist significantly increased to 18.0 ± 1.98 polyps, an overall 85.7% polyp detection rate (P < .001). Simulated polyp detection rates were also significantly higher by using the ultrawide-viewing mode in each colon segment, 4.5 ± 0.65 polyps (90.0%) right colon, 4.0 ± 0.87 (80.0%) polyps transverse colon, and 9.6 ± 1.28 polyps (87.3%) left colon (all comparisons, P < .001). Importantly, the ultrawide-viewing mode detected significantly more "hidden" simulated polyps (81.9% vs 31.9%, P < .0001).

Limitations: Nonrandomized design, use of a colon model, and "simulated" colon polyps.

Conclusions: Ultrawide-view colonoscopy significantly improved simulated polyp detection in a colon model. Clinical studies in human subjects should be pursued to further evaluate this new endoscopic technology.

Publication types

  • Comparative Study
  • Multicenter Study
  • Video-Audio Media

MeSH terms

  • Adult
  • Attitude of Health Personnel
  • Clinical Competence
  • Colonic Polyps / diagnosis*
  • Colonoscopes*
  • Colonoscopy / instrumentation
  • Colonoscopy / methods*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Models, Anatomic
  • Optical Devices
  • Prospective Studies