Background: Breast cancer sentinel lymph node (SLN) biopsy has become a common procedure. The GeneSearch™ Breast Lymph Node Assay is a real-time reverse-transcriptase polymerase chain reaction assay for detecting nodal metastases larger than 0.2 mm. The trial is a prospective multi-center clinical trial conducted to validate the assay in China.
Methods: Four hundred and seventy-nine consecutive prospective patients were enrolled from six centers. SLNs were sectioned along the short axis into multiple blocks. Odd blocks were tested by the assay intra-operatively, and even blocks were assessed by post-operative histology. Six 4- to 6-μm-thick sections were taken every 150 μm per block. In addition, intra-operative histological assessments were performed on the even blocks of 214 patients by frozen section (FS) and all blocks of 156 patients by touch imprint cytology (TIC).
Results: A total of 1046 SLNs were excised. Overall performance of the assay compared to post-operative histology was accuracy of 91.4 %, sensitivity of 87.5 %, and specificity of 92.9 %. There were no significant differences in assay performance of each center. After a learning curve of about 10 cases, the assay could be performed in a median time of about 35 min. The sensitivity of the assay was similar to the FS (84.9 %, P = 0.885) and was significantly higher than the TIC (70.0 %, P = 0.007) while the specificity of all were comparable.
Conclusion: The GeneSearch™ Breast Lymph Node Assay is an accurate and rapid intra-operative assay for breast SLNs and it can replace FS and TIC for application.