The paradigm shift of globalized research and conducting clinical studies at different geographic locations worldwide to access broader patient populations has resulted in increased need of correlating bioanalytical results generated in multiple laboratories, often across national borders. Cross-validations of bioanalytical methods are often implemented to assure the equivalency of the bioanalytical results is demonstrated. Regulatory agencies, such as the US FDA and European Medicines Agency, have included the requirement of cross-validations in their respective bioanalytical validation guidance and guidelines. While those documents provide high-level expectations, the detailed implementation is at the discretion of each individual organization. At Bristol-Myers Squibb, we practice a fit-for-purpose approach for conducting cross-validations for small-molecule bioanalytical methods using LC-MS/MS. A step-by-step proposal on the overall strategy, procedures and technical details for conducting a successful cross-validation is presented herein. A case study utilizing the proposed cross-validation approach to rule out method variability as the potential cause for high variance observed in PK studies is also presented.