Improved encoding pulses for Bloch-Siegert B1(+) mapping

J Magn Reson. 2013 Jan:226:79-87. doi: 10.1016/j.jmr.2012.11.004. Epub 2012 Nov 17.

Abstract

A new family of optimized encoding pulses for Bloch-Siegert (BS) |B(1)(+)| mapping is introduced, as well as an algorithm to design them. The pulses are designed by numerical maximization of BS sequence sensitivity, subject to constraints that ensure low on-resonance excitation. The pulses are in all cases characterized by a constant envelope and U-shaped frequency sweep. They are validated in simulations, 7T in vivo experiments, and an experiment to measure their on-resonance excitation, and are compared to a Fermi pulse conventionally used in the BS method. The pulses are shown to produce larger phase shifts in a shorter time and with lower on-resonance excitation than the Fermi pulse, which results in lower SAR and improved |B(1)(+)| accuracy in areas of the body with large main field inhomogeneities.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Algorithms*
  • Brain Mapping / methods
  • Computer Simulation
  • Electromagnetic Fields
  • Humans
  • Image Enhancement
  • Magnetic Resonance Imaging / methods*
  • Magnetic Resonance Imaging / statistics & numerical data*
  • Phantoms, Imaging
  • Reproducibility of Results