Prediction of fluid responsiveness by a continuous non-invasive assessment of arterial pressure in critically ill patients: comparison with four other dynamic indices

Br J Anaesth. 2012 Sep;109(3):330-8. doi: 10.1093/bja/aes182. Epub 2012 Jun 26.

Abstract

Background: We evaluated the ability of an infrared photoplethysmography arterial waveform (continuous non-invasive arterial pressure, CNAP) to estimate arterial pulse pressure variation (PPV). We compared the ability of non-invasive PPV to predict fluid responsiveness with invasive PPV, respiratory variation of pulse contour-derived stroke volume, and changes in cardiac index induced by passive leg raising (PLR) and end-expiratory occlusion (EEO) tests.

Methods: We measured the responses of cardiac index (PiCCO) to 500 ml of saline in 47 critically ill patients with haemodynamic failure. Before fluid administration, we recorded non-invasive and invasive PPVs, stroke volume variation, and changes in cardiac index induced by PLR and by 15 s EEO. Logistic regressions were performed to investigate the advantage of combining invasive PPV, stroke volume variation, PLR, and EEO when predicting fluid responsiveness.

Results: In eight patients, CNAP could not record arterial pressure. In the 39 remaining patients, fluid increased cardiac index by ≥15% in 17 'responders'. Considering the 195 pairs of measurements, the bias (sd) between invasive and non-invasive PPVs was -0.6 (2.3)%. The areas under the receiver operating characteristic (ROC) curves for predicting fluid responsiveness were 0.89 (95% confidence interval, 0.78-1.01) for non-invasive PPV compared with 0.89 (0.77-1.01), 0.84 (0.70-0.96), 0.95 (0.88-1.03), and 0.97 (0.91-1.03) for invasive pulse pressure, stroke volume variations, PLR, and EEO tests (no significant difference). Combining multiple tests did not significantly improve the area under the ROC curves.

Conclusions: Non-invasive assessment of PPV seems valuable in predicting fluid responsiveness.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Blood Pressure*
  • Critical Illness / therapy
  • Fluid Therapy*
  • Humans
  • Logistic Models
  • Middle Aged
  • Photoplethysmography
  • ROC Curve
  • Stroke Volume