Mitral isthmus ablation using steerable sheath and high ablation power: a single center experience

J Cardiovasc Electrophysiol. 2012 Nov;23(11):1193-200. doi: 10.1111/j.1540-8167.2012.02380.x. Epub 2012 Jun 15.

Abstract

Background: Mitral isthmus ablation is challenging. The use of steerable sheath and high ablation power may improve success rate.

Methods: This single-center, prospective study enrolled 200 patients who underwent ablation for atrial fibrillation (AF), including mitral isthmus ablation. Mitral isthmus ablation was performed using an irrigated ablation catheter via a steerable sheath (endocardium: maximum power: 40/50 W limited to annular end, maximum temperature: 48 °C; coronary sinus [CS]: maximum power: 25/30 W, maximum temperature: 48 °C). Endpoint was bidirectional mitral isthmus block.

Results: Mitral isthmus block was acutely achieved in 182/200 patients (91%). Sixty-nine percent of patients required CS ablation. Mean total ablation time was 13 ± 6 minutes. There was 1 case of acute circumflex artery occlusion. Mean left atrium (LA) diameter was significantly bigger in patients with unsuccessful mitral isthmus ablation (49 ± 4 mm vs. 43 ± 6 mm; P = 0.0007). In redo procedures, the incidence of reconduction at the mitral isthmus, roof and cavotricuspid isthmus was 44%, 37%, and 29%, respectively. Overall incidence of perimitral flutter was 9%. Prior complex fractionated atrial electrogram ablation was a predictor for microreentrant atrial tachycardia (AT) whereas gaps in linear lesions predicted macroreentrant flutters. After a mean follow-up of 20 ± 9 months, 73% of patients remained free from AF or AT.

Conclusion: We reported on a series of mitral isthmus ablation using steerable sheath and high ablation power (50 W). Larger LA diameter was a predictor of failure to achieve mitral isthmus block. The mitral isthmus had a moderately high incidence of re-conduction but was only associated with a relatively low incidence of perimitral flutter.

MeSH terms

  • Aged
  • Atrial Fibrillation / diagnosis
  • Atrial Fibrillation / physiopathology
  • Atrial Fibrillation / surgery*
  • Atrial Flutter / etiology
  • Cardiac Catheters*
  • Catheter Ablation / adverse effects
  • Catheter Ablation / instrumentation*
  • Catheter Ablation / methods*
  • Chi-Square Distribution
  • Electrocardiography
  • Electrophysiologic Techniques, Cardiac
  • England
  • Equipment Design
  • Female
  • Humans
  • Logistic Models
  • Male
  • Middle Aged
  • Mitral Valve / physiopathology
  • Mitral Valve / surgery*
  • Multivariate Analysis
  • Prospective Studies
  • Recurrence
  • Reoperation
  • Tachycardia, Supraventricular / etiology
  • Therapeutic Irrigation / instrumentation
  • Time Factors
  • Treatment Outcome