Suture-mediated arterial access site closure after transfemoral aortic valve implantation

Catheter Cardiovasc Interv. 2013 Feb;81(2):E139-50. doi: 10.1002/ccd.24326. Epub 2012 May 2.

Abstract

Objectives: To evaluate feasibility, safety and efficacy of percutaneous arterial access site closure after transfemoral, transcatheter aortic valve implantation (TF-TAVI) using a single, commercially available six French monofilament suture-mediated vascular closure device (VCD) in "preclosure" technique.

Background: Currently, TF-TAVI is evolving into a completely percutaneous procedure. However, percutaneous access site closure still remains a major technical challenge with room for improvement.

Methods: 94 of 144 consecutive patients underwent completely percutaneous TF-TAVI using following technique for access site closure: After puncture of the common femoral artery using fluoroscopy and contralateral angiography for guidance, the VCD was deployed prior further predilatation of the vessel and insertion of the large-bore introducer sheath. At the end of the procedure, the preloaded sutures were tied for final hemostasis and crossover angiography was used for postprocedural evaluation of the access vessel.

Results: Application of the VCD was technically successful in all cases, resulting in an efficient hemostasis with cessation of any bleeding within 10 min of final knot-tying in 83 of the 94 patients, and there was only one closure-failure with continuous bleeding despite prolonged manual compression requiring endovascular treatment. In addition, we observed four closure-related access vessel stenoses of hemodynamic relevance requiring endovascular treatment in three and surgical repair in one patient. However, interventional and surgical repair was not associated with death or irreversible end-organ damage and all patients recovered without sequelae.

Conclusion: "Preclosure" of the arterial access site with a single six French suture-mediated VCD is relatively easy, safe and efficient method for access site closure after TF-TAVI which, along with ongoing profile reductions of TAVI devices, should further simplify and broaden the way toward a routine, completely percutaneous procedure.

Publication types

  • Evaluation Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Analysis of Variance
  • Aortic Valve Stenosis / diagnosis
  • Aortic Valve Stenosis / therapy*
  • Arterial Occlusive Diseases / etiology
  • Arterial Occlusive Diseases / physiopathology
  • Arterial Occlusive Diseases / therapy
  • Cardiac Catheterization* / adverse effects
  • Chi-Square Distribution
  • Endovascular Procedures
  • Feasibility Studies
  • Femoral Artery / diagnostic imaging
  • Femoral Artery / surgery*
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / methods*
  • Hemodynamics
  • Hemorrhage / etiology
  • Hemorrhage / prevention & control*
  • Hemostatic Techniques* / adverse effects
  • Humans
  • Punctures
  • Radiography, Interventional
  • Retrospective Studies
  • Severity of Illness Index
  • Suture Techniques* / adverse effects
  • Treatment Outcome