Trivalent and hexavalent chromium have been successfully separated and determined using low pressure ion chromatography combined with flow injection spectrophotometric analysis (LPIC-FIA). A column packed with crosslinking starch microspheres was used for on-line separation of Cr(III) from Cr(VI) in a flow-injection system because of its absorptive effect on Cr(III). To determine the concentration of Cr(III) and Cr(VI) in samples, we used 3.0 mmol/L nitric acid to elute adsorbed Cr(III) from the column and then used ceric sulfate-sulfuric acid as oxidant to convert all Cr(III) into Cr(VI). Then, Cr(VI) directly came from the samples and Cr(VI) came from Cr(III) successively formed a amaranthine complex with diphenycarbazide and the complex shows a maximum absorption at 530 nm. Analytical parameters including the concentration of eluent and oxidant solution, oxidizing temperature, length of oxidizing reaction coil, reaction coil and injection coil, interfering effects, etc., were optimized. The limit of detection was 1.25 μg/L for Cr(VI) and 3.76 μg/L for Cr(III). The linear relationship between absorption with the concentration of Cr(VI) and Cr(III) was 0.001-1.000 mg/L and 0.030-1.000 mg/L with correlation coefficients of 0.9995 and 0.9994, respectively. The relative standard deviation of Cr(VI) and Cr(III) was 1.21% and 1.66%, respectively (n=10). Major cations and anions did not show any interference. We validated this method through certified reference materials and through measuring the recovery in tannery wastewater.