A novel in situ probe for oxygen uptake rate measurement in mammalian cell cultures

Biotechnol Prog. 2012 Mar-Apr;28(2):581-6. doi: 10.1002/btpr.741. Epub 2011 Nov 14.

Abstract

The newly developed in situ oxygen uptake rate (in situ OUR) probe presented in this article is based on the in situ microscope technology platform. It is designed to measure the oxygen uptake rate (OUR) of mammalian cells, an important parameter for metabolic flux analysis, inside a reactor (in situ) and in real-time. The system isolates a known volume of cell culture from the bulk inside the bioreactor, monitors the oxygen consumption over time, and releases the sample again. The sample is mixed during the measurement with a new agitation system to keep the cells in suspension and prevent oxygen concentration gradients. The OUR measurement system also doubles as a standard dissolved oxygen (DO) probe for process monitoring when it is not performing OUR measurements. It can be equipped with two different types of optical sensors (i.e., DO, pH) simultaneously or a conventional polarographic DO-probe (Clark type). This new probe was successfully tested in baby hamster kidney perfusion cell cultures.

Publication types

  • Evaluation Study

MeSH terms

  • Animals
  • Cell Culture Techniques
  • Cells / chemistry
  • Cells / metabolism*
  • Cricetinae
  • Kinetics
  • Molecular Probe Techniques*
  • Molecular Probes / chemistry
  • Oxygen / chemistry
  • Oxygen / metabolism*

Substances

  • Molecular Probes
  • Oxygen