The three year follow-up of the randomised "all-comers" trial of a biodegradable polymer biolimus-eluting stent versus permanent polymer sirolimus-eluting stent (LEADERS)

EuroIntervention. 2011 Nov;7(7):789-95. doi: 10.4244/EIJV7I7A125.

Abstract

Aims: The current study reports clinical outcomes at three year follow-up of the LEADERS clinical trial which was the first all-comers trial comparing a new generation biodegradable polymer biolimus drug-eluting stent (BES) with the first generation permanent polymer sirolimus-eluting stent (SES).

Methods and results: One thousand seven hundred and seven patients were randomised to unrestricted use of BES (n=857) or SES (n=850) in an all-comers population. Three year follow-up was available in 95% of the patients, 812 treated with BES and 809 treated with SES. At three years, BES remains non-inferior to SES for the primary endpoint of major adverse cardiac events (composite of cardiac death, myocardial infarction (MI), or clinically-indicated target vessel revascularisation (CI-TVR) (BES 15.7% versus SES 19%; HR 0.82 CI 0.65-1.03; p=0.09). The MACE Kaplan Meier event curves increasingly diverge with the difference in events increasing from 1.4% to 2.4% and 3.3% at 1, 2 and 3 years, respectively in favour of BES. The rate of cardiac death was non-significantly lower 4.2% versus 5.2% (HR=0.81 CI 0.52-1.26; p=0.34) and the rate of myocardial infarction was equivalent 7.2% versus 7.1% (HR 1.01 CI 0.70-1.44; p=0.97) for BES versus SES, respectively. Thus BES was non-inferior to SES in all the safety endpoints. Clinically-indicated TVR occurred in 9.4% of BES treated patients versus 11.1% of SES treated patients (HR 0.84 CI 0.62-1.13; p=0.25). Rates of definite stent thrombosis were 2.2% for BES and 2.9% for SES (HR 0.78 CI 0.43-1.43; p=0.43), with the event rate increase of 0.2% from one to three years for BES and 0.9% for SES. For patients presenting with ST-elevation myocardial infarction BES was superior to SES in reducing MACE.

Conclusions: The findings of the three year follow-up support the claim that the biodegradable polymer biolimus-eluting stent has equivalent safety and efficacy to permanent polymer sirolimus-eluting stent in an all-comers patient population. Its performance is superior in some subpopulations such as in ST-elevation MI patients and event rates for BES are overall lower than for SES with a trend toward increasing divergence of outcomes over three years.

Publication types

  • Comparative Study
  • Multicenter Study
  • Randomized Controlled Trial

MeSH terms

  • Absorbable Implants*
  • Aged
  • Angioplasty, Balloon, Coronary / adverse effects
  • Angioplasty, Balloon, Coronary / instrumentation*
  • Angioplasty, Balloon, Coronary / mortality
  • Cardiovascular Agents / administration & dosage*
  • Chi-Square Distribution
  • Coronary Artery Disease / mortality
  • Coronary Artery Disease / therapy*
  • Drug Therapy, Combination
  • Drug-Eluting Stents*
  • Europe
  • Female
  • Humans
  • Kaplan-Meier Estimate
  • Male
  • Middle Aged
  • Myocardial Infarction / etiology
  • Myocardial Infarction / prevention & control
  • Platelet Aggregation Inhibitors / therapeutic use
  • Polymers*
  • Proportional Hazards Models
  • Prosthesis Design
  • Risk Assessment
  • Risk Factors
  • Sirolimus / administration & dosage
  • Sirolimus / analogs & derivatives*
  • Thrombosis / etiology
  • Thrombosis / prevention & control
  • Time Factors
  • Treatment Outcome

Substances

  • Cardiovascular Agents
  • Platelet Aggregation Inhibitors
  • Polymers
  • umirolimus
  • Sirolimus