Transmit beamforming for optimal second-harmonic generation

IEEE Trans Ultrason Ferroelectr Freq Control. 2011 Aug;58(8):1559-69. doi: 10.1109/TUFFC.2011.1983.

Abstract

A simulation study of transmit ultrasound beams from several transducer configurations is conducted to compare second-harmonic imaging at 3.5 MHz and 11 MHz. Second- harmonic generation and the ability to suppress near field echoes are compared. Each transducer configuration is defined by a chosen f-number and focal depth, and the transmit pressure is estimated to not exceed a mechanical index of 1.2. The medium resembles homogeneous muscle tissue with nonlinear elasticity and power-law attenuation. To improve computational efficiency, the KZK equation is utilized, and all transducers are circular-symmetric. Previous literature shows that second-harmonic generation is proportional to the square of the transmit pressure, and that transducer configurations with different transmit frequencies, but equal aperture and focal depth in terms of wavelengths, generate identical second-harmonic fields in terms of shape. Results verify this for a medium with attenuation f1. For attenuation f1.1, deviations are found, and the high frequency subsequently performs worse than the low frequency. The results suggest that high frequencies are less able to suppress near-field echoes in the presence of a heterogeneous body wall than low frequencies.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Acoustics
  • Algorithms
  • Body Constitution
  • Computer Simulation
  • Humans
  • Image Processing, Computer-Assisted / methods*
  • Models, Theoretical*
  • Transducers
  • Ultrasonography*