Background and aim: Palliative biliary decompression by metal stent is the treatment of choice for unresectable malignant biliary obstruction; however, conventional stents provide only mechanical palliation and exert no anti-tumor effects. Gemcitabine (GEM) has been reported to be more effective in unresectable pancreatic cancer and biliary cancer compared with other chemotherapeutic drugs. We evaluated the safety of a GEM-eluting stent by analyzing histologic responses of the porcine bile duct.
Methods: Stents containing GEM (0%, 10%, 15%, and 20% [w/v]) were surgically inserted into bile ducts of pigs (each group, n = 2). The animals were euthanized after 4 weeks, and the stented bile duct segment underwent gross and microscopic examination. Laboratory assay was performed for aspartate transaminase (AST), alanine transaminase (ALT), total bilirubin, and gamma-glutamyl transferase (γ-GTP).
Results: Moderate to severe inflammation was observed in the bile ducts in contact with stents containing 15 and 20% GEM, compared with no inflammation with 0% GEM and mild inflammation with 10% GEM. Fibrous reactions observed in the submucosal layer did not differ among groups. Transmural necrosis and perforations were not observed in any animal. No abnormal laboratory test findings were directly caused by GEM.
Conclusion: Our newly developed GEM eluting stents can be used safely in normal bile ducts. Our results indicated that 10% GEM produced mild histologic changes in the stented segment and adjacent tissue; this concentration may be appropriate for clinical application.