A prospective randomized FDA-IDE trial comparing Cortoss with PMMA for vertebroplasty: a comparative effectiveness research study with 24-month follow-up

Spine (Phila Pa 1976). 2012 Apr 1;37(7):544-50. doi: 10.1097/BRS.0b013e31822ba50b.

Abstract

Study design: Randomized, controlled, prospective (AAOS therapeutic level I) trial conducted under Food and Drug Administration Investigational Device Exemption hypothesized noninferiority of a novel bioactive composite material to polymethylmethacrylate (PMMA).

Objective: To determine the safety and efficacy of a new, nonresorbable bioactive composite (Cortoss) compared with PMMA, the standard treatment of vertebral compression fractures (VCFs).

Summary of background data: Vertebroplasty with PMMA, the widely adopted treatment for VCFs nonresponsive to conservative care, provides effective, immediate pain relief, but the material has received criticism for its properties. A new bioactive composite material designed specifically for vertebroplasty showed promising results in animals and nonrandomized clinical studies and was subsequently compared with PMMA in a randomized study. This study represents the first prospective study evaluating vertebroplasty comparing PMMA with another material with 24-month follow-up.

Methods: Using 2:1 randomization, vertebroplasty was performed with Cortoss on 162 patients and PMMA on 94 patients. Inclusion criteria were painful osteoporotic VCFs with a visual analogue scale pain score of at least 50 mm on a 100-mm scale and at least 30% disability as measured by the Oswestry Disability Index. Evaluations at pretreatment, treatment day, and 7 posttreatment intervals out to 24 months included pain, disability, neurological status, adverse events, quality of life, patient satisfaction, analgesic use, and independently reviewed radiographs.

Results: Noninferiority of Cortoss relative to PMMA was observed, with Cortoss-treated patients experiencing significant pain relief at 3 months (P = 0.0395) and better maintenance or improvement in function at 24 months (P = 0.0299). Incidence of serious device-related adverse events was 4.3% in both groups; none were life threatening.

Conclusion: Vertebroplasty using either Cortoss or PMMA provides effective, immediate, and lasting pain relief and prevents further loss of function. Both materials performed comparably at most time points; Cortoss demonstrated better results for pain reduction at 3 months and for function at 24 months.

Publication types

  • Comparative Study
  • Multicenter Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Bisphenol A-Glycidyl Methacrylate / therapeutic use*
  • Comparative Effectiveness Research
  • Female
  • Follow-Up Studies
  • Fractures, Compression / surgery*
  • Humans
  • Male
  • Middle Aged
  • Pain Measurement
  • Polymethyl Methacrylate / therapeutic use*
  • Prospective Studies
  • Spinal Fractures / surgery*
  • Treatment Outcome
  • Vertebroplasty / instrumentation*
  • Vertebroplasty / methods

Substances

  • CORTOSS
  • Bisphenol A-Glycidyl Methacrylate
  • Polymethyl Methacrylate