The Sequenom MassARRAY iPLEX single-nucleotide polymorphism (SNP) typing platform uses matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) coupled with single-base extension PCR for high-throughput multiplex SNP detection. In this study, we investigated the use of iPLEX MassARRAY technology for methicillin-resistant Staphylococcus aureus (MRSA) genotyping. A 16-plex MassARRAY iPLEX GOLD assay (MRSA-iPLEX) was developed that targets a set of informative SNPs and binary genes for MRSA characterization. The method was evaluated with 147 MRSA isolates, and the results were compared with those of an established SYBR Green-based real-time PCR system utilizing the same SNP-binary markers. A total of 2352 markers belonging to 44 SNP-binary profiles were analysed by both real-time PCR and MRSA-iPLEX. With real-time PCR as the reference standard, MRSA-iPLEX correctly assigned 2298 of the 2352 (97.7%) markers. Sequence variation in the MRSA-iPLEX primer targets accounted for the majority of MRSA-iPLEX erroneous results, highlighting the importance of primer target selection. MRSA-iPLEX provided optimal throughput for MRSA genotyping, and was, on a reagent basis, more cost-effective than the real-time PCR methods. The 16-plex MRSA-iPLEX is a suitable alternative to SYBR Green-based real-time PCR typing of major sequence types and clonal complexes of MRSA.
© 2011 The Authors. Clinical Microbiology and Infection © 2011 European Society of Clinical Microbiology and Infectious Diseases.