We report a novel microfluidic chamber incorporating fluid ports with active suction to achieve localized chemical stimulation of brain slices. A two-level soft-lithography process is used to fabricate fluid ports with integrated injection and suction holes that are connected to underlying microchannels. Fluorescence imaging, particle tracking velocimetry, and cell staining are used to characterize flows around a fluid port with or without active suction to validate effective localization of injected chemicals. To demonstrate biological applicability of the chamber, we show an induction of cortical spreading depression (CSD) waves in mouse brain slices through controlled focal delivery of potassium chloride solution.
This journal is © The Royal Society of Chemistry 2011