Background and purpose: In some patients, angiographic flow impairment is observed during carotid artery stenting (CAS) using Angioguard XP (AGXP), resulting in neurological symptoms. CAS was thus modified to improve clinical outcome.
Methods: Ninety-seven patients were treated with CAS using AGXP from January 2008 to October 2009. In period I (January-December 2008; n = 53), blood aspirations were performed only in no-flow cases. In period II (January-October 2009; n = 44), blood aspirations were performed in no-flow and slow-flow cases. Clinical outcome, detection of microembolic lesions on diffusion-weighted imaging (DWI) and flow impairment during CAS were examined between these two periods before and after modifying the CAS procedure.
Results: Periprocedural transient ischemic attacks occurred in 10 patients (18.9%) and one patient (2.27%) in periods I and II, respectively (P = .018). Minor and major strokes were observed in two patients in each period (P = .849). New ipsilateral DWI lesions were detected in 25 patients (47.2%) and 11 patients (25.0%) in periods I and II, respectively (P = .024). Among 18 slow-flow cases, new DWI lesions were detected in one patient (9.09%) and five patients (71.4%) with (n = 11) and without (n = 7) blood aspirations, respectively (P = .013). Neurological symptoms were observed only in three of seven patients (42.9%) without aspirations, compared to one of 11 patients (9.1%) with aspirations (P = .043).
Conclusion: Postoperative symptomatic stroke and new DWI lesions are significantly associated with blood flow impairment during CAS using AGXP. When flow impairment occurs, blood aspiration should be performed.