Highly sensitive fluorescence detection system for microfluidic lab-on-a-chip

Lab Chip. 2011 May 7;11(9):1664-70. doi: 10.1039/c0lc00586j. Epub 2011 Mar 24.

Abstract

We demonstrate a compact, low cost and practical fluorescence detection system for lab-on-a-chip applications. The system comprises a commercially available InGaN light emitting diode (501 nm) as light source, an organic or silicon photodiode detector, absorptive dye coated colour filters and linear and reflective polarisers. An injection moulded polystyrene microfluidic chip is used as the platform for fluorescence immunoassays for cardiac markers myoglobin and CK-MB. The optical limit of detection (LOD) is measured using a TransFluoSphere® suspension at 5.6 × 10(4) beads µl(-1) which can be equated to ∼3 nM fluorescein equivalent concentration. The LOD for the human plasma immunoassays is measured as 1.5 ng ml(-1) for both myoglobin and CK-MB.

MeSH terms

  • Biomarkers / blood
  • Creatine Kinase, MB Form / blood
  • Fluorescent Antibody Technique / methods*
  • Humans
  • Lab-On-A-Chip Devices*
  • Microfluidic Analytical Techniques / methods*
  • Myoglobin / blood
  • Point-of-Care Systems
  • Sensitivity and Specificity

Substances

  • Biomarkers
  • Myoglobin
  • Creatine Kinase, MB Form